OPERATING PROCEDURES

Figure 15 shows the TB82 keypad and display with all possible icons in
the display activated. For any given operation only the icons involved in
the operation will be shown.
Key Functions
Exit to MEASURE - Escapes back to the Measure Mode from all modes or
programming states of operation.
FAULT Info - Accesses information on diagnostic or error conditions.
SELECT - Selects the mode or programming state of operation shown in the
secondary display region
ENTER - Accepts the mode or programming state of operation shown in the
secondary display region
NEXT - Increments through a series of programming states
YES - Affirms the action that is about to take place
NO - Denies the action that is about to take place.
MENU - Increments through the modes of operation
Increases numerals & moves through a series of parameters
Moves the flashing data entry value one space to the right
Decreases numeric values & moves through a series of parameters
Figure 16 shows the TB82 in the MEASURE mode of operation.
Pressing the key at the far right, underneath the MENU icon, provides
access to all instrument functions.
Figure 17 shows the display after the MENU key has been pressed
once. The CALIBRATE menu can now be accessed by pressing the key
underneath the SELECT icon.
display will change to the next menu function choice, OUT/HOLD.
Pressing again will activate the CONFIGURE menu choice, and so on.
There are six MENU functions:
MEASURE - Normal operational display
CALIBRATE – Allows calibration of the conductivity and temperature sensors,
restoring calibration to the original factory values, tuning of the analog
output and also allows the ability to view or change the sensor's slope and
offset.
OUT/HOLD - Allows the output to be re-ranged, held or released (Analog version
only)
CONFIGURE - Sets up how the transmitter operates
SECURITY - Sets password protection to the CONFIGURE, CALIBRATE and
OUTPUT menus
DISPLAY - Displays temperature in °F or °C, current output in mA, sensor input,
sensor group, concentration spike output status and software revision. Sets
what will be displayed under the secondary display when in the measure
mode.
To commission the TB82:
1.
Configure the instrument as required for the application.
2.
Calibrate the sensor(s) as required (TB82EC and TB82TC).
3.
Set up parameters on the receiving devices (e.g. PLC, DCS recorder) to properly recognize the TB82 outputs.
TB82COND-ATEX-EN-A
TB82 Conductivity Series ATEX Product Instruction Manual
If the MENU key is pressed again the

Commissioning the Instrument

Hidden Key
ABB
(User Mode)
OUTPUT
SPIKE
FAULT
HELD
300
%
Mohms-cm
mS/cm
μS/cm
mS/m
25°C
M
exit to
MEASURE
SELECT
YES
FAULT
ENTER
NEXT
info
Figure 15
300
μS/cm
25°C
M
Figure 16
300
μS/cm
Calibr
exit to
MEASURE
SELECT
Figure 17
TB82
MEASURE
CALIBRATE
ppm
ppb
OUT/HOLD
S/m
CONFIGURE
SECURITY
DISPLAY
NO
MENU
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
MENU
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
MENU
13