FORCE (MOD. FREQUENCY)
ENG
When?
Why?
How?
Example
Electrodes
Intensity
Option 2+2
FORCE (MOD. FREQUENCY) (30 MIN)
Frequency
Duration of ramp-up
Duration of phase
Duration of ramp-down
11
To maintain muscular strength during immobilization, or to increase
muscular strength for a specific muscle/muscle group.
The programme focuses on the Type II fibres to improve stamina in the
muscle.
By modulating the frequency for a period of time that is equal to the
contraction time, the contraction feels more comfortable.
Increased strength for hip abductors. Stimulation on Quadriceps, in case
of immobilization.
Electrodes positioned as indicated according to the stimulated muscle.
See suggested electrodes placements on the fold out cover.
Maximum tolerable stimulation energy, which is one of the key factors
for the effectiveness of the treatment. The higher the stimulation
energy, the greater the number of muscular fibres (motor units) working.
Try to gradually increase the energy level during a treatment session.
If the patient experiences discomfort and pain during stimulation, the
2+2 function allows this programme (ch 1+2) to be combined with a pain
relief programme (ch 3+4).
Warm-up
6 Hz
1.5 s
2 min
2 s
Contraction
Active rest
35-60 Hz
4 Hz
3 s
0.5 s
8 s
15 s
1 s
0.5 s
Final recovery
3 Hz
1.5 s
3 min
3 s