Disinfection
Please carefully observe the points below when disin-
fecting your atomiser and accessories. We recommend
disinfecting the individual parts on a daily basis after the
last usage as a minimum measure.
(All you need for this is a little white vinegar and distilled
water.)
• First, clean the atomiser and accessories as described
in the "Cleaning" section.
• Place the disassembled atomiser [7], the mouthpiece
[11], the silicone masks [12, 13] and the disassembled
nosepiece [n/a] in boiling water for 5 minutes.
• For the remaining accessories, use a vinegar solution
consisting of ¼ vinegar and ¾ distilled water. Make
sure that the volume is sufficient to fully submerse
parts, such as the PVC masks [n/a] and hose [6], in
the solution. Accessories that can be boiled can also
be disinfected in this way.
• Leave the parts in the vinegar solution for 30 minutes.
• Rinse the parts with water and dry them carefully with
a soft cloth.
Important
Do not boil or autoclave the compressed air hose or the
PVC masks [n/a].
• Put the parts together again when they are completely
dry and place them in a dry, sealed container.
Note
• Please ensure that the parts are completely dried af-
ter cleaning, otherwise the risk of bacterial growth is
increased.
• Use a cold disinfection solution in accordance with the
manufacturer's instructions.
Drying
• Place the individual parts on a clean, dry and absor-
bent surface and leave them to dry completely (at least
4 hours).
Durability of materials
• As with any plastic parts, atomisers and their acces-
sories are affected by a certain amount of wear and
tear when used and hygienically prepared on a fre-
quent basis. Over time, this can lead to a change in
the aerosol, which can have a negative effect on the
efficiency of the treatment. We therefore recommend
that you replace the atomiser and other accessories
after a year.
• The following point should be observed when selecting
the cleaning product or disinfectant: only use a mild
cleaning product or disinfectant in the concentration
and volume prescribed by the manufacturer.
Storage
• Do not store in damp conditions (such as in a bath-
room) and do not transport with any damp items.
• When storing and transporting, protect from prolonged
direct sunlight.
• The accessories can be stowed securely in the ac-
cessory compartment. Store the device in a dry place,
ideally in the original packaging.

9. Disposal

For environmental reasons, do not dispose of the
device in the household waste.
Please dispose of the device in accordance with
EC Directive – WEEE (Waste Electrical and Electronic
Equipment).
If you have any questions, please contact the local au-
thorities responsible for waste disposal.