Description
1 Locking nuts
2 Adjustment spindle
3 Collet
4 Hexagon wrench (in the handle)
5 Guard
6 Upper anvil
7 Lower anvil
8 Backstop
9 Fine setting screw
10 Liftoff-lever (only on model 301 P)
5
3
1
6
8
7
Illustration shows the 301 P
Preparations
Prior to using a Mahr dial indicator it is required to screw
in the supplied spherical contact point (length 20 mm).
- Insert the indicator into the snap gage. At the nominal
size the upper anvil (6) should have an advance of
0,15 to 0,2 mm.
- Tighten up the collet (3) with the hexagon wrench 5/8"
resp. 16 mm.
- Clean the backstop (8) and both anvils (6 & 7) with a
soft cloth.
- Loosen the locking nuts (1) by using the hexagon
wrench (4) as lever.
- Push the backstop (8) all the way in to the end of the
guide.
Setting to nominal size
- Insert the master into the the snap gage so that it rests
across the centre of the lower anvil (7).
- Finger tighten the locking nuts (1) so that a slight
resistance is felt during the adjustment.
- Using the hexagon wrench (4) as lever set the
adjustment spindle (2) such that the pointer of the
indicator is roughly at the zero position.
- Turn back the adjustment spindle (2) about 1/4 turn.
- Set the backstop (8) such that the workpiece rests
across the centre of the lower anvil (7).
- Using the hexagon wrench (4) as lever tighten up the
locking nuts (1).
- Set the pointer of the indicator with the fine setting
screw (9) to zero.
GB
USA
4
9
2
10
Anzeigende Rachenlehre
Indicating Snap Gage
300 P/301 P
Bedienungsanleitung
Operating Instructions
3756690
Mahr GmbH Esslingen
D-73702 Esslingen • Postfach 10 02 54
Telefon (07 11) 9 31 26 00 • Fax (07 11) 3 16 09 53
0702
loading

Inhaltszusammenfassung für Mahr 300 P

  • Seite 1 Illustration shows the 301 P 0702 Preparations Prior to using a Mahr dial indicator it is required to screw in the supplied spherical contact point (length 20 mm). - Insert the indicator into the snap gage. At the nominal size the upper anvil (6) should have an advance of 0,15 to 0,2 mm.
  • Seite 2 Basic equipment consists of: Zur Grundausstattung gehören: – Indicating Snap Gage 300 P or 301 P – Anzeigende Rachenlehre 300 P bzw. 301 P – Key for hexagon socket screw –...

Diese Anleitung auch für:

301 p