Environment
• Observe the surface temperature on the
product, before placing an user in it. This
goes especially for users with insensitive
skin as they cannot feel the heat. If the
surface temperature is above 41°C then
allow the product to cool before use
• For the purpose of safety and comfort,
always check the water temperature before
bathing the child
Product
• Carry out all positional adjustments on the
product and accessories and ensure that
all knobs, screws and buckles are securely
fastened before use. Keep all tools out of
the reach of children
• Ensure the stability of the product is
maintained before placing the user in it
Service information
Maintenance of a medical device is wholly the
responsibility of the owner of that device. Failure
to maintain a device in accordance with the
instructions may invalidate the device's warranty.
Furthermore, failure to maintain a device may
compromise the clinical condition or safety of
users and/or their caregivers.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
Before every use
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Check all fittings and buckles for signs of
worn out parts
After every use
• Disinfect the product. please check the
disinfection section in this User Guide for
more information
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
support.R82.org
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Warewashing machine
This product can be washed at 60° using a mild
detergent for 10 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
This product can be washed at 85° using a mild
detergent for 3 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• Suction cups are clean and able to ensure
firm attachment to the bottom of a bathtub
when placing the product
• The pommel is adjustable and can be fixed
in position
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
11
EN