Preparation for

Manual Cleaning

1
Cleaning was validated using
Enzol® at 1 oz/gal. at 35-40°C.
Manual Cleaning
2
Cleaning was validated using
Renu-Klenz™ at ¼ oz/gal. at
35-40°C.
Disassemble the case into its individual
1
components: the base, lid, and (if applicable)
silicon mat.
Prepare an enzymatic detergent
2
the manufacturer's recommendations.
Wipe the entire case with a clean cloth dipped
3
in the detergent.
Immerse the case in the detergent.
4
Soak the case in the detergent for at least
5
15 minutes.
Brush
1
• Thoroughly brush the exterior of the case with
a soft-bristled brush, focusing on any mated or
rough surfaces.
• Using a pipe cleaner and syringe, brush and
flush the hard-to-reach areas of the case.
• Actuate the case, brushing any movable parts
in their extreme open and closed positions.
Rinse
2
• Rinse the case with reverse osmosis/de-
ionized (RO/DI) water at ambient temperature
to remove all detergent residues. After all
detergent residues have been removed,
continue to rinse for a minimum of 30 seconds.
• Drain excess water from the case and dry it
using a clean cloth and pressurized air (40 psi).
• Visually inspect the case for cleanliness, paying
close attention to hard-to-reach areas. If visible
soil remains, repeat steps 1 and 2.
Soak
3
• Prepare a non-enzymatic detergent
to the manufacturer's recommendations.
• Fully immerse the case and inject any mated
surfaces with at least 50mL of the detergent.
• Soak the case for a minimum of 15 minutes.
according to
1
according
2
5

Werbung