Herunterladen Diese Seite drucken

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29

Introduction

This reprocessing guide provides instructions for the proper cleaning and sterilization
of the following sterilization case.
233-032-130
This sterilization case is intended for use only with the devices listed below. For case/
device configuration, see the "Sterilization Case Setup" section of this guide.
Part Number
233-070-001
275-023-027
275-035-302
With any 11 of the below instruments
Part Number
233-071-001
233-071-002
233-071-003
233-071-004
233-071-005
233-071-006
233-071-007
233-071-008
233-071-011
233-071-013
233-071-014
With any two of the below instruments.
Part Number
275-035-100
275-035-200
These reprocessing instructions are provided in accordance with ISO 17664, AAMI
TIR 12, AAMI ST81. While they have been validated by Stryker as being capable of
preparing the device for re-use, it remains the responsibility of the processor to
ensure that the reprocessing as actually performed, using equipment, materials,
and personnel in the reprocessing facility, achieves the desired result. This normally
requires validation and routine monitoring of the process. Stryker recommends users
observe these standards when reprocessing medical devices.
2
Small-Joint Rigid Instrument Sterilization Case
Desciption
Camelback Handle
2.7mm Scope Bridge
3.5mm Small Joint Cutter Cannula
Desciption
2.0mm Angled Probe
1.5mm Straight Probe
2.0mm Straight Rasp
2.0mm Back Angled Rasp
2.0mm Up Angled Rasp
2.0mm Retrograde Knife
2.5mm Hook Knife
2.0mm Smillie Knife
2.0mm Serrated Straight Knife
1.5mm Straight Scalpel Knife
1.5mm Curved Scalpel Knife
Desciption
3.5mm Trocar
3.5mm Obturator

Werbung