Herunterladen Diese Seite drucken

Stryker Surgical Anleitung Seite 10

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
Automated Cleaning
6
Brush
4
• Thoroughly brush the exterior of the case using
a soft-bristled brush.
• Inject the prepared detergent into any mated
surfaces a minimum of five times.
• Actuate the case, brushing any movable parts
in their extreme open and closed positions.
• Using a pipe cleaner and syringe, brush and
flush the hard-to-reach areas of the case.
Rinse
5
• Thoroughly rinse the case with ambient RO/
DI water until all detergent residue is removed.
Flush any crevices five times with a syringe.
After the detergent residue is removed,
continue rinsing for a minimum of 30 seconds.
• Drain the excess water from the case and dry it
with a clean cloth and pressurized air (40 psi).
• Visually inspect the case for cleanliness, paying
close attention to hard-to-reach areas.
Rinse
1
• Disassemble the case into its individual
components: the base, lid, and (if applicable)
silicon mat.
• Rinse the case with RO/DI water at ambient
temperature until there is no visible soil. Use
a syringe to assist in rinsing. After all visible
soil has been removed, continue to rinse for a
minimum of 30 seconds.
• Place the case in the washer on an incline to
facilitate drainage.

Werbung

loading