Herunterladen Diese Seite drucken

Stryker Surgical Anleitung Seite 13

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
Temperature
Relative Humidity
Vacuum Set Points
EtO Concentration
Time
Wrapping
Temperature
Time
Dry Time
Ethylene oxide (EtO)
Preconditioning
55°C
(131°F )
70%
1.30 psia
60 minutes
Autoclave (Steam)
Gravity
double
132°C
(270°F)
30 minutes
45 minutes
Warning: Drying time depends on several
variables, including: altitude, humidity, type of
wrap, preconditioning, size of chamber, mass
of load, material of load, and placement in
chamber. Users must verify that drying time set
in their autoclave yields dry surgical equipment.
Warning: These instructions were developed
using the guidance from AAMI TIR 12, ISO 17665
and AAMI ST79 and Stryker recommends users
observe these standards.
Sterilization
55°C
(131°F )
70%
600 mg/L
100% EtO
2 hours
Prevacuum
double
132°C
(270°F)
5 minutes
30 minutes
Aeration
51–59°C
(119–138°F)
12 hours
"Flash"
Prevacuum
132°C
(270°F)
5 minutes
9