Disinfecting the Internal Cavity
Perform the following procedure if you suspect that the Adapter Cable's internal cavity has become
contaminated:
Caution
Always clean and rinse the Adapter Cable's internal cavity before disinfecting (reference the previous
procedure, Cleaning the Internal Cavity).
WARNING
During the following procedure, always point the Adapter Cable's Spiral Electrode Input
from your face, to avoid being sprayed by the discharged disinfectant and rinse.
Step
1
Fill a 10cc syringe with a solution of recommended disinfectant.
Lift open the cover on the Adapter Cable Flush-Cleaning Port
2
3
Hold the Adapter Cable over a sink, then:
A. Insert the syringe into the Flush-Cleaning Port
inject the disinfectant into the cavity.
B. Allow the disinfectant to discharge from the Spiral Electrode
Input
into the sink.
4
Fill the syringe with clean water and repeat steps 3A and 3B to rinse the cavity. Repeat the clean
water rinse injection 2-3 times to assure a thorough rinse.
Using an empty syringe, inject air into the Flush-Cleaning Port
5
excess water. Close the Flush-Cleaning Port
Disinfecting the Internal Cavity (Optional, as Needed)
.
and slowly
. Repeat 4-5 times to remove
before returning the Adapter Cable to use.
6
away
S
lowly inject
disinfectant