Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo InPro series Bedienungsanleitung Seite 24

Optischer sauerstoffsensor
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
en
24
3.2
Simulation of Condition Modes
Select via the navigation key "Condition Mode"
the condition to be simulated.
Condition
SIM O2 ZERO
SIM O2 AIR1
SIM O2 AIR2
SIM O2 TGL
(30 sec)
SIM O2 ERR1
SIM O2 ERR2 0% air LED off
3.3
Simulation of Warnings and Alarms
Simulation of "Signal below"
1. Select the condition mode "SIM O2 ERR1".
Result: The display of the transmitter shows "...
% air". The menu "Messages" shows "Chx Sig-
nal error".
Simulation of "Temperature above the predefined
temperature"
1. Set the LED Mode to "Auto" at the transmitter.
2. Select the condition mode "SIM O2 ERR2".
Result: The display of the transmitters shows
"0.0" and "LED Off". The menu "Messages
shows "Chx LED Off".
Optical O
Simulator
2
30 422 249
InPro6860 i
InPro6870 i
InPro6960 i
0% air 25 °C
100% air 40 °C
50% air 40 °C
0% air 25 °C
100% air 40 °C
...% air 25 °C
65 °C
© 02 / 1 8 Mettler-Toledo GmbH
Optical O
2
InPro6970 i
THO ODO
0% air 5 °C
100% air 25 °C
50% air 25 °C
0% air 5 °C
100% air 25 °C
...% air 25 °C
0% air LED off
65 °C
Printed in Switzerland