OPERATION MANUAL AND INSTRUCTIONS
ANTI-CLOCKWISE ONLY CALIBRATING / TESTING (Fig. 2-b)
Torque tighten to 20 Nm. (or hand tighten bolts with hex key provided)
CLOCKWISE AND ANTI-CLOCKWISE CALIBRATING/ TESTING (Fig. 2-c)
IMPORTANT!
Bolts must be torque tightened to 50 Nm. Replace bolts if removed.
IMPORTANT!
The whole of the transducer must always be in contact with the supporting surface. (Fig. 3)
Locating display housing
Locate Dynatester Display housing on to the bolt heads securing the transducer housing (Fig. 4).
Connecting power supply
Plug Power Supply into back of display housing then plug power cable into Power Supply.
TIP. If the power cable has no plug fitted, wire as follows:
BROWN-LIVE
BLUE-NEUTRAL
GREEN / YELLOW-EARTH
If in doubt consult a qualified electrician.
Switching ON
Switch ON Dynatester ('I' in the down position).
Allow dynatester to equalise to the ambient temperature/humidity before switching on. wipe off
any moisture before use.
TIP.
No display when power is switched on.
a) Check the Power Supply output connector is correctly plugged into the mating part on the
back of the display housing.
b) Check the fuse in the power cable plug (if fitted).
Mode
TRACK
TRACK
TRACK
Use
Allow 5 minutes for the instrument
to warm up and stabilise, then press
ZERO
ZERO
TRACK
How it works
Follows signal
ZERO
EN
Visual representation
19