air. The drain valve (7) drains condensate from the air tank when drying performance is checked.
Compressed, oil-free filtered air is stored in the air tank ready for use.
The pressure vessel does not need to be drained.
Compressor cabinet
The soundproof cabinet provides compact covering of the compressor, allowing sufficient exchange
of cooling air. With its design It can be placed in a dentist's office as a part of furniture. The fan
under the air pump of a compressor provides cooling of compressor and it is in operation at the
same time with an engine of the compressor. After prolonged use the temperature in the cabinet
may rise above 40°C, causing the cooling fan blower to automatically turn on. After cooling the
cabinet area down to 32°C the fan blower turns off automatically.
Make sure that nothing impedes the free flow of air under and around the
compressor. Never cover the hot air outlet on the top back side of the cabinet.
If placing the compressor on a soft floor such as carpet, create space for ventilation
between the base and floor or the cabinet and floor, e.g. underpin the footings with
hard pads.
The 8-10 bar compressor model is equipped with an hour counter (Fig. 3).
Fig. 1 - DK50-10 Z - Compressor
NP-DK50-10-7_05-2019-MD
- 14 -
Description for Figures 1-3
1. Air pump
2. Air tank
3. Non-return valve
4. Pressure switch
5. Safety valve
6. Pressure gauge
7. Drain valve
8. Inlet filter
9. Dryer
10. -
11. -
12. -
13. Solenoid valve
14. Dryer cooler
15. Filter
16. Condensate drain
solenoid valve
17. Compressed air outlet
18. Circuit breaker switch
19. Bottle
20. Magnetic holder
21. Fan
22. Switch
23. Hour counter
24. Socket
25. Cabinet fan
DK50-10
05/2019

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Dk50-10 sDk50-10 z/mDk50-10 s/m

Inhaltsverzeichnis