Herunterladen Diese Seite drucken

Connection Of The Fore-Vacuum Pump - Agilent Technologies TwisTorr 84 FS X3502-64000 Bedienungshandbuch

Werbung

Connection of the Fore-Vacuum Pump

A flange KF 16 NW is available to connect the TwisTorr 84 FS pump
to the forevacuum pump. A hose or vacuum approved pipe can be
used. If a rigid pipe is used, any vibration generated by the
mechanical pump must be eliminated through the use of bellows.
NOTE
The TwisTorr 84 FS pump is characterized by its high compression ratio also for
oil vapors. When using a mechanical oil sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore vacuum pump in order to
prevent oil back streaming.
TwisTorr 84 FS User Manual / 87-901-043-01(B)
Technical Information

Connection of the Fore-Vacuum Pump

17
325/336

Werbung

loading