Herunterladen Diese Seite drucken

Connection C - Electrical - Agilent Technologies TwisTorr 84 FS X3502-64000 Bedienungshandbuch

Werbung

17
Technical Information
Interconnections
Connection C – Electrical
The turbopump is connected to the controller through a 6-pin
connector. Pins B, C and D are the 3-phase supply to the motor, pins
A and F are connected to the temperature sensor (NTC type, 30 KΩ
resistance at 25 °C) and pin E is connected to the pump ground.
If the temperature sensor is disconnected, the pump will not start. To
prevent damage to the pump when the temperature exceeds 60 °C,
the sensor automatically cuts off the power supply.
302/336
TwisTorr 84 FS User Manual / 87-901-043-01(B)