TROUBLESHOOTING

The compressor must not be repaired except by the Technical
Service of PARI GmbH or a service center expressly
authorised to do so by PARI GmbH. If the compressor is
opened or manipulated by anyone else, all claims under the
warranty shall be void. In these cases, PARI GmbH will accept
no liability.
Fault
The com-
pressor does
not start up.
No aerosol is
coming out of
the nebuliser.
– 33 –
Possible cause
The power plug is
not plugged into the
mains socket
properly.
The supply voltage is
not suitable for
running the
compressor.
The nebuliser nozzle
insert is blocked.
The compressed air
supply tube is not
connected correctly
to the compressed
air connection on the
compressor or the
nebuliser lower
section.
The tube attached to
the compressed air
connection on the
compressor is
leaking.
Remedy
Check whether the power
plug is plugged into the
socket correctly.
Check whether the local
supply voltage matches the
voltage data marked on the
compressor identification
label.
Clean the nebuliser.
Check that the tubing
connectors are connected
firmly to the compressor and
the nebuliser.
Replace the defective tube
or the entire PARI SINUS
tubing system.