• Stop the measurement.
• Select the program branch Parameter. Press ENTER until FLUID c-Calibr. is displayed.
FLUID c-Calibr.
Edit? no
>YES<
c-Calibr.
1
TYPE
c-Calibr.
[m/s]
0.0 K0
...
c-Calibr.
[m/s]
Save? no
>YES<
Concentr.Correc.
0.000
M%
• Repeat the steps if it is necessary to correct K
Offset
An offset is only valid for the operating point and should, therefore, only be entered after the calibration coefficient K
been determined exactly.
Concentr.Correc.
0.000
M%
• Create a start-up protocol (see annex A).
Further optimization
If a further optimization of the measurement is necessary, send the stored measurement data, the start-up protocol, the ref-
erence values and additional information about the process conditions during the measurement to FLEXIM (see annex A).