Supplement to User Manual - ENGLISH
Output Test
4
mA
Ix= 4.0mA
Again? >NO<
yes
If the transmitter has more than one current output, repeat the steps for the other list items.
1.5

Selection of the Physical Quantity and the Units of Measurement

After all the current outputs have been installed, the physical quantity and the unit of measurement will be set.
par mea>OPT<sf
Output Options
Output Options
for Channel
A:
Physic. Quant.
Volume flow
Volume in:
m3/h
...
Unit of measurement of the temperature and the density
SYSTEM settings ↕
Dialogs/Menus
...
Temperature
>[°C]<
[°F]
...
Density [lb/ft3]
no
>YES<
Density unit
g/cm3
>kg/m3<
...
Unit of measurement of the concentration
The unit of measirement of the concentration is determined by the FluxKoef file that is being used.
SUPIOX-SV3-0, 2011-11-29
Press ENTER.
Select no. Press ENTER.
Select the program branch Output Options. Press ENTER.
Select the measuring channel to which the flow transducers are connected. Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Select the physical quantity. Press ENTER.
Concentration may not be selected as the physical quantity.
If actual measure is selected for an output, the list item that is to be used for this output
has to be selected as the physical quantity (see section 1.4).
Select the unit of measurement of the selected physical quantity. Press ENTER.
This display will not be indicated if the sound speed is selected as the physical quantity.
Enter the output options.
Select Special Funct.\SYSTEM settings\Dialogs/Menus.
Press ENTER until Temperature is displayed.
Select the unit of measurement of the temperature.
Press ENTER until Density [lb/ft3] is displayed.
3
Select yes if lb/ft
is to be used or no. Press ENTER.
Select the unit of measurement of the density. Press ENTER.
This display will only be indicated if no is selected.
Press ENTER until SYSTEM settings\Dialogs/Menus is displayed again. The set-
tings are stored.
Measurement with PIOX S
25