Supplement to User Manual - ENGLISH
Store Concentr.
off
>ON<
par mea>OPT<sf
Output Options
Store Meas.Data
no
>YES<
Storage Rate
Once per 10 sec.
Storage Rate
1
s
1.8

Measurement

• Select the measuring point for the flow transducers (see user manual UMFLUXUS_F7, chapter "Selection of the
Measuring Point").
• Start the measurement (see user manual UMFLUXUS_F7).
S=■■■■■■
Q=■■■■■■■■■■■■
K=Concentration
A:K= 82.4
M%
K=Concentration
A:K=FAIL_xx
M%
Tab. 1.1: Error codes of the concentration measurement
error code
error
01
invalid coefficients or algorithm
02
numerical error during the calculation
04
temperature outside of the range of the coefficients
08
concentration outside of the range of the coefficients
10
medium temperature not available
20
sound speed not available
If several errors occur, the error code will be the sum of the hexadecimal numbers, e.g. 0C = 04 + 08.
SUPIOX-SV3-0, 2011-11-29
Select on to store the concentration. Press ENTER.
Select the program branch Output Options.
Press ENTER until Store Meas.Data is displayed. Select yes to activate storing. Press
ENTER.
Select a storage rate or EXTRA. Press ENTER.
recommended value during the start-up: 1...10 s
recommended value during operation: 1...10 min
This display will only be indicated if Store Meas.Data and/or Serial Output are acti-
vated.
If EXTRA is selected, enter the storage rate. Press ENTER.
Check if the signal amplitude S and the signal quality Q are sufficient.
The concentration (in mass percent) can be displayed.
If the concentration can not be determined, an error message will be displayed, with xx
being a bit-coded hexadecimal error code (see Tab. 1.1).
Measurement with PIOX S
27