Measurement with PIOX S
1.1.2
Mass Flow Measurement (Optional)
If the transmitter has the option Mass flow, the mass flow rate is available as an additional physical quantity.
To calculate the operating density of the medium in dependence on the concentration and the temperature, an algorithm is
used that has been defined in the program FluxKoef and transmitted to the transmitter:
ρ = f(K, T)
The mass flow rate is calculated from the operating density and the volumetric flow rate.:
·
.
m ·
= ρ
V
with
ρ
-
operating density
K
-
concentration
T
-
temperature
m ·
-
mass flow rate
·
V
-
volumetric flow rate
1.2

Activation of the Measurement with PIOX S

Measurement of the concentration or the mass flow rate (optional) has to be activated:
SYSTEM settings↕
Measuring
Enable Concentr.
no
>YES<
1.3

Installation of Temperature Measurement

The temperature is measured by:
• connecting a temperature probe to the temperature input of the transmitter or
• connecting an external temperature measurement to the current input of the transmitter
FLEXIM Temperature Probe Pt100, Pt1000
The temperature probe has to be installed close to the flow transducers in order to obtain temperature values at the mea-
suring point.
• Connect the temperature probe to the temperature input T1 of the transmitter.
• Select the type of the temperature probe:
SYSTEM settings ↕
Proc. inputs
Proc. inputs
PT100 / PT1000
Input T1
>PT100<
pt1000
• Install the temperature input:
SYSTEM settings ↕
Proc. inputs
Proc. inputs
Link temperature
A:T-Inlet
No measuring
22
Select Special Funct.\SYSTEM settings\Measuring. Press ENTER.
Press ENTER until Enable Concentr. is displayed. Select yes. Press ENTER.
Select Special Funct.\SYSTEM settings\Proc. inputs. Press ENTER.
Select the list item PT100 / PT1000.
Select the type.
Select Special Funct.\SYSTEM settings\
Proc. inputs. Press ENTER.
Select Link temperature. Press ENTER.
Select No measuring. Press ENTER.
Supplement to User Manual - ENGLISH