Supplement to User Manual - ENGLISH
A:T-Fluid/Outle ↕
Input T1
A:T(3)
↕
No measuring
A:T(4)
↕
No measuring
SYSTEM settings ↕
Proc. inputs
External Temperature Measurement
Make sure that the measured temperatures correspond to the conditions at the measuring point where the flow transduc-
ers are mounted.
• Connect the current output of the external temperature measurement to the current input I1 of the transmitter.
• Install the temperature input:
SYSTEM settings ↕
Proc. inputs
Proc. inputs
↕
Define ext.meas.
Ext.meas.value
↕
From input I1
I1:Input for
↕
Temperature
I1:Sensor MIN
4.00
mA
I1:Sensor MAX
20.00
mA
I1:MIN(4.0mA)
↕
0.00
C
I1:MAX(4.0mA)
↕
100.00
C
Ext.meas.value
↕
...go back
Proc. inputs
↕
Link temperature
A:T-Inlet
↕
No measuring
SUPIOX-SV3-0, 2011-11-29
Select Input T1. Press ENTER.
Select No measuring. Press ENTER.
Select No measuring. Press ENTER.
Press ENTER until the scroll list SYSTEM settings is displayed again.
Select Special Funct.\SYSTEM settings\Proc. inputs. Press ENTER.
Select Define ext.meas. Press ENTER.
Select From input I1. Press ENTER.
Select Temperature. Press ENTER.
Enter the min. value of the external current output. Press ENTER.
Enter the max. value of the external current output. Press ENTER.
Enter the temperature that corresponds to the min. value of the external current output.
Press ENTER.
Enter the temperature that corresponds to the max. value of the external current output.
Press ENTER.
Select go back. Press ENTER.
Select Link temperature. Press ENTER.
Select No measuring. Press ENTER.
Measurement with PIOX S
23