Measurement with PIOX S
A:T-Fluid/Outle ↕
Input I1
A:T(3)
↕
No measuring
A:T(4)
↕
No measuring
1.4
Installation of the Current Output
SYSTEM settings ↕
Proc. outputs
Install Output
↕
Current Ix
(√)
Ix enable
no
>YES<
Ix disable
>NO<
yes
Ix Source chan. ↕
Channel A:
Source item
↕
Measuring value
Measuring value ↕
Flow
Ix Output range ↕
4/20 mA
...
Error-value
↕
Other value...
...
24
Select Input I1. Press ENTER.
Select No measuring. Press ENTER.
Select No measuring. Press ENTER.
Select Special Funct.\SYSTEM settings\Proc. outputs. Press ENTER.
Select Current Ix. Press ENTER.
This display will be indicated if the output has not been installed yet. Select yes. Press
ENTER.
This display will be indicated if the output has already been installed. Select no to change
its configuration. Press ENTER.
Select in the scroll list the measuring channel to be assigned to the output as the source
channel. Press ENTER.
This display will not be indicated if the transmitter has only one measuring channel.
Select the source item. Press ENTER.
output
concentration
volumetric flow rate
temperature
mass flow rate (optional)
density (optional)
Select the list item. Press ENTER.
output
concentration
volumetric flow rate
temperature
mass flow rate (optional)
density (optional)
Select the output range. Press ENTER.
Enter the values of the output range, if requested. Press ENTER.
Select an error value. Press ENTER.
Enter the error value, if requested. Press ENTER.
source item
Miscellaneous
Measuring value
Temperature
Measuring value
Miscellaneous
list item
Concentration K
actual measure
(not Flow!)
Tfluid
actual measure
Density
Supplement to User Manual - ENGLISH
SUPIOX-SV3-0, 2011-11-29