5.0 Sterilization:
5.1 Perforator must be manually cleaned prior to sterilization. See
Paragraph 4.0 for recommended procedure.
5.2 Full Cycle Steam Guidelines are as follows:
Cycle #
Cycle Type
1
Pre-Vac,
wrapped
2
Gravity,
wrapped
5.3 The use of other sterilization methods is outside of ACRA-CUT's
responsibility. If the described equipment, materials and chemicals
are not available, it is the responsibility of the user to validate their
method accordingly.
5.4 Frequent reprocessing has few effects on these instruments. The end
of the instrument's working life is normally determined by wear and
damage through use.
www.acracut.com
2-3
Warm as
delivered from
hot water tap
As
N/A
required
As
Ambient
required
Temp (C°)
134
121
Rinse thoroughly
for required time
immediately after step 2.
All surfaces should be
wiped with lint free cloth
soaked in disinfectant or
should be sprayed with a
disinfectant spray.
Allow to air dry in clean
area. Blow lumens with
clean air using filtered air
source or syringe.
Exposure Time
(Minutes)
4
30
English
Dry Time
(Minutes)
20
30
5