Diode Testing; Frequency Measurement; Capacitance Measurement - Monacor DMT-3030 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Diode testing

Set the rotary switch to the position
placing the test tips at the diode in forward direction,
the diode forward voltage is directly displayed in the
4 V range. Here the positive pole of the test voltage
is at the red measuring cable.
With open diode or polarity in reverse direction,
the open circuit voltage of approx. 3 V used with this
measurement is displayed. In case of diode short-
circuit 0 V is displayed.

Frequency measurement

Set the rotary switch to the Hz position, in 5 ranges,
with auto range mode only, the frequency of an AC
voltage up to 500 kHz can be displayed. The input
voltage necessary for this is 2.5 to 3.5 V peak AC
voltage.
The multimeter is protected up to 250 DC/peak
AC voltage.

Capacitance measurement

. When
Set the rotary switch to position
range mode or manual range selection capacitors
from approx. 4 nF up to 40 µF can be measured.
The capacitors to be measured must be dis-
charged.
For values below 40 nF also the relative measure-
ment is recommended. (See below).
The multimeter is protected up to 250 V DC/peak
AC voltage.
The special function
of relative measurement
Display of a measuring value deviating from a
measuring value which has been fixed before.
Example: 10 V have been measured and fixed as a
reference. A following measurement of 11 V is now
displayed as 1 V, i. e. +1 V higher then the reference
. In the auto