6. Commissioning, operation

6. Commissioning, operation

6.1 Mechanical mounting

Only use original accessories.
6.1.1 Sealing the process connection
WARNING!
Risk of burns!
The cooling element can be hot as a result of previous use. Allow the cooling
element to cool beforehand.
The sealing faces at the instrument have to be undamaged and clean.
Parallel threads
per EN 837
Correct sealing of the process connections
with parallel threads at the sealing face 
must be made using suitable flat gaskets,
sealing rings or tecsis profile sealings.
Operating instructions pressure sensor P3278
per DIN 3852-E
Tapered threads
NPT, R and PT
For sealing process connections with
tapered threads, the sealing must be made
in the threads using additional sealing
material, e.g. PTFE tape (EN 837-2).