Herunterladen Inhalt Inhalt Diese Seite drucken

Simatic Pdm - Siemens SITRANS LR200 Kurzanleitung

Profibus pa
Vorschau ausblenden Andere Handbücher für SITRANS LR200:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 46
1.5. Low Calibration Point
Distance from Sensor Reference to Low Calibration Point: (usually process empty level).
Range: 0.0000 to 20.000 m
Values
1.6. High Calibration Point
Distance from Sensor Reference to High Calibration Point (usually process full level).
Range: 0.0000 to 20.000 m
Values
1.7. Apply? (Apply changes)
In order to save the Quick Start settings it is necessary to select Yes to apply changes.
YES, NO, DONE (display shows DONE when Quick Start is successfully com-
Options
pleted.)
Press Mode
to return to Measurement mode. SITRANS LR200 is now ready to operate.
Requesting an Echo Profile
1)
In PROGRAM mode, navigate to: Level Meter >
Diagnostics (3) > Echo Profile (3.2).
2)
Press RIGHT arrow
3)
In the Profile screen, press UP arrow
the Transmit icon, and RIGHT arrow
the profile.
4)
Press DOWN arrow
then RIGHT arrow
SITRANS LR200 Communications: PROFIBUS PA
• You will need the full manual to acquire the list of applicable parameters.
• We recommend that you use SIMATIC Process Device Manager (PDM) to program your
device.

SIMATIC PDM

Notes:
• SIMATIC PDM Rev. 6.0 SP3 or higher is recommended.
• SIMATIC PDM Rev. 5.2 SP1 is supported only for basic configuration and troubleshooting.
For advanced features such as the Quick Start wizard, Revision 6.0 SP3 or higher is
required.
SIMATIC PDM is a software package used to commission and maintain SITRANS LR200 and
other process devices. Please consult the operating instructions or online help for details on
using SIMATIC PDM. You can find more information at www.siemens.com/simatic-pdm.
Page EN-14
SITRANS LR200 (PROFIBUS PA) – QUICK START MANUAL
to request a profile.
to select
to update
to select the Exit icon,
to return to previous menu.
DIAGNOSTICS
3.2
FAULT RESET
ECHO PROFILE
MEAS. VALUES
REMAIN DEV. LIFE
transmit icon,
deselected
C:38 A:TF D:4.25
exit icon,
selected
1
2
3
4
5
transmit icon,
selected
7ML19985XD81
exit icon,
deselected

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis