Press RIGHT arrow
values and move directly to the next item.
d)
To change a setting, scroll to the desired item or key in a new value.
e)
After modifying a value, press RIGHT arrow
to move to the next item.
f)
Quick Start settings take effect only after you select Yes to Apply changes in
step 1.7.
1.1. Material
Options
1.2. Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Options
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
1.3. Sensor Units
Selects the units for the Quick Start variables (high and
low calibration point, and level, distance, or space).
1.4. Operation
distance
space
level
NO
SERVICE
LEVEL
Operation
types
SPACE
DISTANCE
7ML19985XD81
SITRANS LR200 (PROFIBUS PA) – QUICK START MANUAL
to open Edit mode or DOWN arrow
LIQUID
LIQUID LOW DK (low dielectric liquid)
SLOW
0.1 m/minute
MED
1.0 m/minute
FAST
10.0 m/minute
sensor
reference
point
high cal.
point
low cal.
point
The SITRANS LR200 stops updating measurements. Last valid
measurement is displayed.
Distance to material surface referenced from Low Calibration
Point (process empty level).
Distance to material surface referenced from High Calibration
Point (process full level).
Distance to material surface referenced from Sensor Reference
Point.
to accept default
to accept it and press DOWN arrow
M, CM, MM, FT, IN
Options
Page EN-13