Herunterladen Diese Seite drucken

Ace instruments ACE X Bedienungsanleitung Seite 20

Atemalkoholtester
Vorschau ausblenden Andere Handbücher für ACE X:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Battery notices
The device associated with this operating manual contains 2 AA Duracell Alkaline batteries.
These batteries are non-rechargeable. The use of a rechargeable battery is not possible with
this device and can damage it. If the included batteries are depleted, they must not be thrown
away into household waste. Used batteries may contain harmful substances that can cause
damage to the environment or your health. Please return all depleted batteries to electrical
appliance stores, who collect them for free.
Battery exchange
To replace the batteries, push the battery cover on the rear of the unit downwards and replace
the discharged batteries with new ones. Be aware of the polarity when inserting the batteries
(+/- pole). Close the battery compartment cover again. In the EU, the return of used batteries is
mandatory by law. All depleted batteries and rechargeable batteries will be recycled. This way
valuable resources like iron, zinc and nickel can be retrieved and reused. Battery recycling is
the easiest lesson in environmental protection – so please support it!
The shown garbage bins mean that batteries or rechargeable batteries must not be thrown
into the household waste. The signs below the garbage bins are acronyms for the chemical
substances inside the battery:
29
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead
Technical data*
Sensor
Premium fuel cell sensor
Accuracy
+/- 0.05‰ (per mill) with reference to 1.00‰ at 25°C
(‰ = default setting)
OR +/- 0.005% BAC (Blood Alcohol Concentration)
with reference to 0.100% BAC at 25°C
OR +/- 0.025mg/L with reference to 0.50mg/L at 25°C
Detection Range
0.00 up to 5.00‰ (‰ = default setting)
OR 0.000 up to 0.500% BAC (USA)
OR 0.00 up to 2.50mg/L (UK)
Warm-up Time
Within 10 seconds at 0.5‰. This may vary depending on the value
of per mill / BAC which was measured
Test memory
Stores the results of up to 100 tests
(with test number, date and time)
Dimension
128 x 60 x 24mm (L x W x H)
Weight
130g (batteries included)
Power Supply
2 pcs. of 1.5V AA Alkaline batteries (power for up to 1,000 tests)
Working
-5°C to 40°C
Temperature
Storage
-5°C to 40°C
Temperature
Connectivity
Frequency range: 2402 to 2480 MHz
(only with optional BT)
Maximum transmission power: 1,6 dBm
Calibration
Default setting: After 12 months or 500 tests (whichever comes
first). The settings can be changed (see chapter "Advanced Menu
Settings"). Please observe the calibration instructions on the
display. You can send the instrument for calibration to the service
address given at www.alkomat.net.
*The technical data are subject to change without prior notice for functional improvements.
30

Werbung

loading