ACE INSTRUMENTS ADVISES
EVEN IF YOU GET A RESULT MEANING „SAFE RANGE", YOU SHOULD NEVER DRIVE AFTER
CONSUMING ALCOHOL OR DO ANY OTHER ACTIVITIES THAT ARE SEVERELY IMPAIRED BY
ALCOHOL CONSUMPTION AND MIGHT ENDANGER YOUR HEALTH AND SAFETY OR THE
HEALTH AND SAFETY OF OTHERS.
Status Indicators on the Display
#82
Total number of all tests performed
(will be reset to 0 after a calibration)
Display
NOTICE
MEANING
BLOW ERROR
The sampling volume or blow
power is not enough to analyze
the breath sample.
LOW BATTERY
The testing cannot be
performed because the
battery is low.
CALIBRATION DUE
The recommended calibration
DATE HAS PASSED
is due. This display may
depend on the time factor
(default setting: 12 months). The
setting can be changed under
"Advanced Menu Settings".
OVER NUMBER
The recommended calibration
OF TEST
is due. This display may
ALLOWED
depend on the number of tests
(default setting: 500 tests). The
setting can be changed under
"Advanced Menu Settings".
23
D365
Remaining days until
Remaining battery
calibration
power
ACTION
Press the power button to perform
a new test.
Replace the integrated batteries
with new batteries.
Contact your local distributor or
customer service center for proper
recalibration service.
Contact your local distributor or
customer service center for proper
recalibration service.
BREATH SAMPLE
WITH EXCESSIVE
SALIVA
OUT OF TEMP
RANGE
NO MEMORY
CAPACITY
AVAILABLE
Indications
Basic Menu Options
How to enter the menu options: During the warm-up process, press the menu button for
2 seconds. You will see these menu options:
TEST RESULT
UNIT
INFORMATION
How to move within the menu options:
Š
To move the cursor up, quickly press the power button.
Š
To move the cursor down, quickly press the menu button.
To enter the selected option, press the power button for 2 seconds.
Š
To save the selected setting, press and hold the power button and menu button
Š
simultaneously for 2 seconds
To return to warm up process, press the menu button for 2 seconds.
Š
Excessive saliva went into the
Wait 20 minutes, and then try again.
sensor during the test.
If this error code is displayed for a
long time, contact the ACE Instru-
ments Service Center
(see www.alkomat.net)
The testing cannot be per-
Repeat the test after the device has
formed because the device
reached a temperature within the
is outside of recommended
operating range (-5°C up to 40°C).
operating temperature.
No further test can be per-
Contact your local distributor or the
formed because the number
ACE Instruments Service Center
of tests has already reached
(see www.alkomat.net).
10,000 and no calibration has
been performed so far
(10,000 = maximum number of
tests possible without a single
calibration).
TEST RESULT-G
ALARM LIMIT
BRIGHTNESS
SOUND
LANGUAGE
PIN CODE
24