TV-V position is used for observing a composite video signal at the frame rate.
TV-H position is used for observing a composite video signal at the line rate.
(18)
Trigger SOURCE switch
To conveniently select the trigger source.
VERT: In case of vertical mode switch is CH 1 which automatically becomes registry source.
In case of vertical mode switch is CH 2 which automatically becomes registry source.
Note:
Measurement of VERT condition will be possible only Time/Div. switch is on 0,5 ms/Div. to 0.2
µs/Div. when it is dual mode. When there is no sign on CH 1 & CH 2 place the vertical mode
switch to dual and set TRIGGER SOURCE switch at VERT sweep may flickers but please be
careful as it is not out of order.
CH 1: When there is sign on CH 1 you may select trigger source CH 1.
LINE position selects a trigger derived from the AC power line. This permits the scope to stabilize
display linearelated components of a signal even they are very small compared to other
components of the signal.
EXT position selects the signal applied to the EXT TRIG IN connector.
(9)
Trigger LEVEL control
To select the trigger-signal amplitude at which triggering occurs. When rotated clockwise, the
trigger point moves toward the positive peak of the trigger signal. When its control is rotated
counter-clockwise, the trigger point moves toward the negative peak of the trigger signal.
(8)
Trigger SLOPE switch
To select the positive or negative slope of the trigger signal for initiating sweep.
(19)
EXT TRIG IN connector
For applying external trigger signal to the trigger circuits.
Caution:
To avoid damage to the oscilloscope, do not apply more than 250 V (DC + Peak AC) between
"EXT TRIG IN" terminal and ground.
2.1.4

Miscellaneous Features

(31)
EXT BLANKING INPUT connector
For applying signal to intersity modulate the CRT. Trace brightness is reduced with a positive
signal, and increased with a negative signal.
(17)
PROBE ADJUST
Provides a fast-rise square wave of precise amplitude for probe adjustment and vertical amplifier
calibration.
(28)
Ground connector
Provides an attachment point for separate ground lead.
2.2
Basic Operationg Procedures
The following paragraphs in this section describe how to operate the unit beginning with the most
elementary operating modes, and progressing to the less frequently-used/or complex modes.
2.2.1

Preliminary Control Settings and Adjustments

Before placing the instrument in use, set up and check the instrument as follows: