Precautions; Line Voltage Selection; Installation And Handling Precautions; Operating Instructions - PeakTech 2035 Bedienungsanleitung

30 mhz analog-oszilloskop
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1.3

Precautions

1.3.1

Line voltage selection

This instrument must be operated with the correct Line Voltage Selector Switch setting and the correct
line fuse for the voltage selected to prevent damage. The instrument operates from either a 98 to 125
volts or a 198 to 250 volt line voltage source. Before line voltage is applied to the instrument, make sure
the Line Voltage Selector is set correctly.
To change the line voltage selection:
1.
Make sure the instrument is disconnected from the power source.
2.
Switch the line voltage selector to the desired position.
3.
Pull out the Line Fuse Holder containing the fuse for overload protection. Replace the fuse in the
holder with the correct fuse from table 1-1 and plug it in.
Line voltage
98 to 125 V AC
198 to 250 V AC
1.3.2

Installation and handling precautions

When placing the unit in Service at your workplace, observe the following precautions for best instrument
performance and longest service life.
Avoid placing this instrument in an extremely hot or cold place. Specifically, don't leave this
instrument in a close car, exposed to sunlight in midsummer, or next to a space heater.
Don't use this instrument immediately after bringing it in from the cold. Allow time for it to warm to
room temperature. Similiary don't move it from a warm place to a very cold place, as condensation
might impair its operation.
Do not expose the instrument to wet or dusty environments.
Do not place liquid-filled containers (such as coffee cups) on top of this instrument. A spill could
seriously
damage the instrument.
Do not use this instrument where it is subject to severe vibration, or strong blows.
Do not place heavy objects on the case, or otherwise block the ventilation holes.
Do not use this oscilloscope in strong magnetic fields, such as near motors.
Do not insert wires, tools, etc. through the ventilation holes.
Do not leave a hot soldering iron near the instrument.
Do not place this scope face down on the ground, or damage to the knobs may result.
Do not use this instrument upright while BNC cables are attached to the rear-panel connectors. This
will damage the cable.
Do not apply voltages in excess of the maximum ratings to the input connectors or probes.
2.

Operating instructions

This section contains the information needed to operate the PeakTech 2020 GN or PeakTech 2030 KT
and utilize it in a variety of basic and advanced measurement procedures. Included are the identification
and
Table 1-1: Line voltage Selection and Fuse ratings
Arrow Mark Position
115
230
Fuse Ratings (250 V)
UL198 G
F 1,25 A
F0.63 A
- 33 -
IEC127
F1.25 A
F0.63 A