11.1.2 Electromagnetic compatibility, interference emission

Electromagnetic compatibility, interference emission
The Kröber O2 is intended for use in an environment as described below. The operator of the
Kröber O2 must make sure that the unit is operated in such an environment.
Immunity test
Discharge of static
electricity acc. to
IEC
61000-4-2
fast transient
electric
interference/bursts
acc. to IEC 61000-
4-4
Surge voltages acc.
to
IEC 61000-4-5
Voltage dips, short-
term interruptions
and fluctuations in
supply voltage
IEC 61000-4-11
Magnetic field at a
supply frequency
(50 Hz) acc. to
IEC 61000-4-8
Regulations and declaration of manufacturer -
IEC 60601 test level
±6 kV contact
discharge
±8 kV Air discharge
±2 kV for mains leads
±1 kV for input/output
leads
±1 kV Push-pull
< 5 % U
T
(>95 % dip in U
) for ½
T
period
40 % U
T
(60 % dip in U
5
) for
T
periods
70 % U
T
(30 % dip in U
25
) for
T
periods
<5 % U
T
(95 % dip in U
5 s
) for
T
3 A/m
Compatibility level
±6 kV contact
discharge
±8 kV Air discharge
±2 kV for mains
leads
±1 kV for input/output
leads
±1 kV Push-pull
limited functionality
Mains failure alarm
Restart of unit
limited functionality
Mains failure alarm
Restart of unit
Kröber O2
Appendix
Electromagnetic
environment - regulations
Floors should be made of
wood or concrete or should
be covered with ceramic
tiles. If the floor is covered
with synthetic materials, the
relative humidity must be at
least 30%.
The quality of the supply
voltage should match the
typical office or hospital
environment.
The quality of the supply
voltage should match the
typical office or hospital
environment.
The quality of the supply
voltage should match the
typical office or hospital
environment.
NOTE:
U
is the a.c. supply voltage
T
before application of the test levels
Magnetic fields at mains
frequency must comply with
the typical values found in
office or hospital
environments.