Fig. 7: Scope of delivery 3
7.2 Choosing the location
Please consider the following when choosing the location:
–
The unit should have 30 cm clearance from any walls, curtains and other large objects
(e.g. cupboards), to ensure unrestricted entry of air through the back of the unit.
–
The unit is air cooled. It must therefore not be placed near heaters, etc. At such a
location there is a risk of overheating.
–
The unit should be operated only in places where it is not impaired by air pollution or
smoke.
NOTE!
The unit can be safely lifted and moved by the carrier handle on top of the unit.
ATTENTION! Ensure a sufficient air supply!
Ensure that the air inlet of the Kröber O2 oxygen concentrator is located in a
well-vented spot.
The Kröber O2 oxygen concentrator must not be operated directly next to or
even stacked with other equipment. Ensure a sufficient distance from the walls,
etc.!
5 Nasal cannula for oxygen 2m
6 Nasal cannula for oxygen 5m
7 Oxygen safety hose 15m
7
8 Stand for humidifier
9 Oxygen safety hose 0.33m
10 Angular connector
11 Hose connector „christmas tree style"
10
11
Kröber O2
Taking into service
69