9.3 Maintenance plan
Maintenance and cleaning work must be performed at regular intervals as specified in the
table below.
Interval
Daily
Daily
(in case of stationary or
ambulant use)
Daily
(when used in homes or
for nursing care)
Weekly
Every 14 days
earlier if necessary
Every 4 weeks
Every 4 weeks
(when used in homes or
for nursing care)
Annually / after 5000
operating hours
With changing patients
With changing patients
With changing patients
with changing patients
with changing patients
Cleaning work
The humidifier must be cleaned and disinfected every
day.
Nasal cannula should be disposed of and replaced
every day.
Clean the nasal cannula.
If the "Set for external humidification" is used, it must
be cleaned and disinfected every week.
The Kröber O2 must be cleaned with a damp cloth and
disinfected after.
Replace the coarse dust filter
Replace the nasal cannula.
Replace the air intake filter
NOTE!
With extremely dirty ambient air the filter must be
changed earlier.
The device must be cleaned with a damp cloth and
disinfected after.
Replace the humidifier.
Replace the coarse dust filter
Replace the air intake filter
If the "Set for external humidification" is used, the
bracket must be cleaned and disinfected.
Kröber O2
Maintenance
85