5 Technical data
Model
Classification acc. to MPG
Operating voltage
Ambient operating conditions
Sound pressure level
Power input
Coarse dust filter
Bacteria filter
Fuses
Interface
Weight
IP-Klassifikation
Warranty
O2-concentration
Status display O2-concentration
(at operating temperature)
max. recommended volume flow
Volume flow (at 0 or 7kPa)
max. output pressure:
Overpressure relief
Kröber O2
IIa
230 V, 50 Hz
Operation temperature: +5 to +40 °C
Storage temperature: -25 to +70 °C
Humidity: 15% - 93 % r.H., non-condensing
Atmospheric pressure: 700 to 1060 mbar
35 dB(A)
360 W
in back of housing
behind service flap
mains side: 2 x TT2,5A H 250 V
internal: 1 x T1,0A L 250 V
USB
19.8 kg
IP 21
30.000 hours, maximum 5 years (see chapter 2.4)
1 to 4 l/min.
95 % - 3 %
4 to 5 l/min.
85 % +/- 3 %
5 to 6 l/min.
75 % +/- 3 %
82% status display
60% deficiency display
6 l/min
1 - 6 l/min acc. to pre-setting
70 kPa
200 kPa (oxygen reservoir)
250 kPa (compressor)
Kröber O2
Technical data