8.3 Status messages and alarms
8.3.1 Status messages
NOTE!
Status messages are NOT accompanied with acoustical warning signals!
Status message /
LC display
Oxygen status
alarm
8.3.2 Alarm priorities
We differentiate between two alarm priorities:
Alarm priority
Medium priority:
Low priority:
NOTE!
The alarm priorities can audibly be differentiated by different alarm sound sequences.
With a higher priority the number of alarm signals per unit of time increases.
80
Description
This oxygen concentrator is equipped with an innovative multi-
function sensor to monitor the oxygen concentration of the
oxygen output.
If this drops below 82%, the oxygen status alarm is triggered.
Countermeasures
– Check humidifier and hoses for leaks
– Inform the customer service.
NOTE!
According to the specifications of the unit, the oxygen
status alarm is an allowed situation and not an error state
for volume flow setting of greater than 5.0 lpm.
Description
Quick countermeasures by the user are required.
Utmost attention of the user is required.