Fig. 17. Oxygen outlet connector
1
Fig. 18. Humidifier Max.-mark
2
3
Fig. 19. Humidifier assembly
1
Fig. 20: Humidifier bracket
4 Connect the 15m-oxygen safety tubing
to the connector.
5 Fill the humidifier up to the top MAX.-
mark (1).
6 Screw the angle adapter (1) to the
1
humidifier.
7 Plug the 15 m safety hose for oxygen
(extension hose) onto the angle
adapter (2).
8 Push the 0.33-m safety hose for
oxygen onto the outlet connector of the
humidifier (3).
9 Stand the humidifier into the bracket
(1).
NOTE!
The bracket has the function of
preventing the humidifier from falling
over. It is therefore recommended to
mount the bracket to a wall, a
cupboard, etc.
Kröber O2
Taking into service
75