1. Remove the stylus and holder (see page 1-11).
2. Remove the break stem using a 2 mm AF key.
3. Use a C spanner to remove the probe's front
cover. This will expose the metal eyelid seal and
the inner diaphragm seal. Remove the metal
eyelid and spring. CAUTION – these may fall
out.
4. Wash inside the probe, using clean coolant.
(DO NOT use sharp metal objects to clean out
debris.)

DIAPHRAGM MAINTENANCE

Front cover
with O ring
Captive link
5. Inspect the diaphragm seal for signs of piercing
or damage. In the event of damage, return the
probe to your supplier for repair, as coolant
entering the probe mechanism could cause the
probe to fail.
6. Refit the spring and metal eyelid (the spring's
widest diameter is against the metal eyelid).
7. Refit the remaining components (see pages
1-10 and 1-11).
C spanner
4 Nm
(2.95 lbf.ft)
2 mm AF
2.6 Nm
(1.92 lbf.ft)
Break stem
1-15