Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Code
Reason
Text mes-
sage
F264
Error during setup
Installation/
Setup error
F265
Sensor no longer carries out
a measurement
Measurement
function dis-
turbed
F267
Sensor cannot start
No executable
sensor soft-
ware

Function check

Code
Reason
Text mes-
sage
C700
A simulation is active
Simulation ac-
tive
Out of speci cation
Code
Reason
Text mes-
sage
S600
Temperature of the process-
ing electronics in the non-
Impermissi-
ble electronics
temperature
S601
Level echo in the close range
not available
S602
Compensation echo super-
imposed by medium
Level with-
in the search
range, com-
pensation
echo
Recti cation
Check for correct mounting
and/or parameter settings
Check probe length
Carry out a reset
Disconnect operating voltage
Exchanging the electronics
Send instrument for repair
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Recti cation
Finish simulation
Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "
causes as well as corrective measures.
Recti cation
Check ambient temperature
Insulate electronics
Use instrument with higher
temperature range
Reduce level
100 % adjustment: Increase
value
Check mounting socket
Remove possible interfering
signals in the close range
Use coaxial probe
100 % adjustment: Increase
value
9 Diagnostics and servicing
DevSpec State in CMD 48
Bit 10 of Byte 0 ... 5
Bit 11 of Byte 0 ... 5
No communication possible
DevSpec
State in CMD 48
"Simulation Active" in "Stand-
" and provides information on
DevSpec State in CMD 48
Bit 8 of Byte 14 ... 24
Bit 9 of Byte 14 ... 24
Bit 10 of Byte 14 ... 24
55