Herunterladen Inhalt Inhalt Diese Seite drucken

Warning Instructions - KaWe EUROLIGHT Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für EUROLIGHT:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 8
adapter, and, if necessary, clean the glass bulb of the new
lamp with alcohol. The lamp (along with the adapter)
must be pushed in fully until it touches the back of the
opening. The glass bulb should be clean and free of
7
grease.
lamp by pulling it out. If necessary, clean the glass bulb of
the new lamp with alcohol. The glass bulb should be clean
and free of grease. The lamp must be pushed in fully until
it touches the back of the opening.
remove the black plastic sleeve from the lamp. Put in the
new lamp and then slip the plastic sleeve back onto the
lamp. The glass bulb should be clean and free of grease.
6
Unsuited use/contraindication of the
products:
Any other than the specified clinical diagnostic use
described here is not in accordance with the intended use
of the unit. The manufacturer is not liable for any resulting
damages. There are no known risks. The user alone bears
the risk.
Further information, maintenance,
storage:
When used and stored properly, these otoscopes will
serve you many years. KaWe otoscopes are only to be used
with KaWe-brand replacement parts and accessories.
14
KaWe PICCOLIGHT® C
Swing the magnifying glass
upwards and grasping the
adapter, pull out the lamp.
Remove the lamp from the
KaWe EUROLIGHT® F.O.,
KaWe COMBILIGHT® F.O.,
KaWe PICCOLIGHT® F.O.
Take off the F.O. otoscope head
from the handle and remove the
6,7
KaWe EUROLIGHT® C,
KaWe COMBILIGHT® C
Pull out the lamp in the direction
indicated by the arrow from the
KaWe EUROLIGHT® head and
Guarantee:
When used under normal circumstances and with
attention to these instructions, we guarantee this product
for two years after the date of purchase. (With the
exception of lamps/rechargeable batteries/batteries.)
Should you need further information, or should your
instrument require repair, please contact your dealer.

Warning instructions:

When cleaning, disinfecting and sterilising reusable ear
funnels, special attention is required. A temperature of
134°C is not to be exceeded in an autoclave. It is
recommended, that the reusable ear funnels be cleaned as
soon as possible after their use on a patient.
KaWe otoscopes are to be used at room temperatures of
not greater than 35° C. KaWe otoscopes, like all
electrically-powered diagnostic instruments, are subject
to special safety measures with regard to their
electromagnetic compatibility. The operation of the KaWe
otoscopes can be influenced by mobile high-frequency
communication devices.
Limitation of reprocessing:
Frequent professional cleaning affects the instruments
only minimally. The length of the working life of these
instruments usually depends on the wear and tear and
damage by use.
Instructions
Workstation: Only place the instruments on a clean and
sufficiently-sterilised surface.
Storage and transport: No special requirements
Manual cleaning: The exterior surface of the instrument
can be cleaned with a damp, soft and lint-free cloth. Never
allow the otoscope to come into contact with fluids and
ensure that liquids do not penetrate the housing. As the
otoscope is not intended for use during operational
procedures, simple cleaning and or disinfection with an
alcohol-based surface disinfecting agent is sufficient. Do
not use abrasive cleaning agents! The unit cannot be
sterilised. The magnifying glass is to only be cleaned with
glass cleaner.
Automatic cleaning: No automatic cleaning possible/
required.

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

PiccolightCombilight

Inhaltsverzeichnis