Answer: Check the connection. If you still do not see a green light flash, fully recharge
the transmitter battery. Test the transmitter with the tester. If you still do not see a green
light flash, contact your local help line or representative for assistance. It may be time to
replace your transmitter.
Storing the devices
Store the transmitter, charger, and tester in a clean, dry location at room temperature.
Although not required, you may store the transmitter on the charger. If the transmitter is
not in use, you must charge the transmitter at least once every 60 days.

Specifications

Biocompatibility
Applied parts
Operating conditions
Storage conditions
Battery life
Transmitter frequency
Declaration of Conformity
Transmitter: Complies with ISO 10993-1 for body contact
Transmitter
Sensor
Transmitter Temperature: +32° to +122°F (0° to 50°C)
Caution: When operating the transmitter on a tester in air
temperatures greater than 41°C (106°F), the temperature of
the transmitter may exceed 43°C (109°F).
Transmitter Relative Humidity: 10% to 95% with no condensa-
tion
Charger Temperature: +50° to +104°F (10° to 40°C)
Charger Relative Humidity: 30% to 75% with no condensation
Transmitter Temperature: -4° to +131°F (-20° to +55°C)
Transmitter Relative Humidity: 10% to 100% with condensation
Charger Temperature: +14° to +122°F (-10° to +50°C)
Charger Relative Humidity: 10% to 95% with no condensation
Transmitter: 14 days of continuous glucose monitoring immedi-
ately following a full charge
Charger: Completes 40 typical charging operations using a new
AAA or LR-03 battery
MMT-7703NA 916.5 megahertz
MMT-7703WW 868.35 megahertz
Medtronic declares that this product is in conformity with the es-
sential requirements of Directive 1999/5/EC on Radio and Tele-
communications Terminal Equipment and Directive 93/42/EEC
on Medical Devices.