5 TAKING A READING

1
Hold the probe by its sleeve
2
To take a reading, bring the probe down onto the surface whilst holding it perpendicular
3
For subsequent readings, lift the probe off and then replace it onto the coated surface
DO
Hold the probe by the probe sleeve
Gently place the probe onto the surface
Allow the sleeve to make contact with the
surface - to improve accuracy
The display will dim if inactive for more than 15 seconds and will go 'black' if inactive for the
period defined in Menu/Setup/Display/Screen Timeout.
The gauge will switch off automatically after 5 minutes of inactivity
- - -
indicates reading outside range of probe

6 CALIBRATING THE GAUGE

1
Press the Cal softkey
2
For alternative calibration methods, select Cal/Cal Method
3
Select Calibrate and follow the on-screen instructions
4
When prompted place the probe on the centre of the foil(s)
Foils can be stacked
Not all calibration methods are available for all gauge types
see 7 Calibration Methods
5
DO NOT
Drag the probe over the coated surface
Bang the probe down hard onto the surface
Allow the probe to hover over the surface as
this could result in a false reading
90°
www.elcometer.com