Herunterladen Inhalt Inhalt Diese Seite drucken

Electromagnetic Immunity / Mobile Radiofrequency Equipment - Satelec P5 Newtron XS Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Voltage dips, short
interruptions and
voltage variations
IEC 61000-4-11.

10. 3 ELECTROMAGNETIC IMMUNITY / MOBILE RADIOFREQUENCY EQUIPMENT

The device is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: -These specifications may not be applicable in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the device
(including its cables) at a distance less than that recommended and calculated according to the
Conducted
disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated
radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an
electromagnetic environment measurement (a), must be less than the compliance level in each
frequency range (b).
Interference may occur near equipment marked with the symbol below:
(a): -The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for
mobile telephones (cellular/cordless), mobile radios, amateur radios, AM/FM radio broadcasts and TV
broadcasts cannot be determined exactly by theory. To assess the electromagnetic environment due to
common mode
common mode
<5% U
(>95% dip
<5% U
T
in U
) for 0.5
in U
T
cycles.
cycles.
40% U
(60% dip
40% U
T
in U
) for 5
in U
T
cycles
cycles
70% U
(30% dip
70% U
T
in U
) for 25
in U
T
cycles
cycles
<5% U
(>95% dip
<5% U
T
in U
) for 250
in U
T
cycles
cycles
IEC
60601
Compliance
test
level
level
frequency and power of the emitter.
3 V/m
150 KHz
3 V/m
to 80
MHz
3 V/m
80 MHz
3 V/m
to 2.5
GHz
(>95% dip
T
) for 0.5
T
Mains power quality should be that of a
(60% dip
T
typical commercial or hospital
) for 5
T
environment.
If the use of the device requires continued
operation during power mains
(30% dip
T
interruption, it is recommended that the
) for 25
T
product be powered from a separate
power supply (UPS, etc.).
(>95% dip
T
) for 250
T
Electromagnetic environment - guidance
Recommended separation distance:
d = 1.2
P
d = 1.2
80MHz to 800MHz.
P
d = 2.3
800MHz to 2.5GHz.
P
Where P is the maximum power rating of the
emitter in watts (W) according to the
manufacturer's specifications and d is the
recommended minimum separation distance in
meters (m).