Herunterladen Inhalt Inhalt Diese Seite drucken

Recommended Separation Distances; Cable Lengths - Satelec P5 Newtron XS Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

fixed radiofrequency emitters, an electromagnetic environment measurement must be made. If the
measured radiofrequency field strength in the immediate environment where the product is used
exceeds the compliance level specified above, the performance of the product must be tested to verify
whether it conforms to the specifications. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the product.
(b): -In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3
V/m.

10. 4 RECOMMENDED SEPARATION DISTANCES

The device is intended for use in an electromagnetic environment in which radiated radiofrequency
disturbances are controlled.
The device user and/or installer can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile radiofrequency communications equipment (emitters)
and the device, according to the maximum output power of the equipment, as recommended in the
table below.
Rated max. power of the emitter
(W)
0.01
0.1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meteres
(m) can be estimated using the equation applicable to the frequency of the emitter, where P is the
max. power rating of the emitter in watts (W) according the manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: -These specifications may not be applicable in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.

10. 5 CABLE LENGTHS

Cables and
accessories
Handpiece cords
Footswitch cord
Power supply cord
Separation distance in metres (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12m
Maximum
length
RF emission, CISPR 1 - Class B/Group 1
Harmonic current emission: IEC61000-3-2
Voltage fluctuation: IEC61000-3-3
Immunity to electrostatic discharge: IEC61000-4-2
Immunity to electrical fast transients/bursts: IEC61000-4-4
Immunity to surges: IEC61000-4-5
Less than 3m
Immunity to voltage dips, short interruptions and voltage
variations: IEC61000-4-11
Immunity to conducted disturbances induced by radiofrequency
fields: IEC61000-4-6
Immunity to radiated radiofrequency electromagnetic fields:
IEC61000-4-3
80 MHz to 800 MHz
d = 1.2
P
0.12 m
0.38 m
1.2 m
3.8 m
12m
Complies with:
800 MHz to 2,5 GHz
d = 2.3
P
0.23 m
0.73 m
2.3 m
7.3 m
23m
P

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

P5 newtron xs led

Inhaltsverzeichnis