- Do not use any cleaning agent in spray form on the device.
a) Maintenance
The device control unit, the handpiece cord and the footswitch must be cleaned and disinfected on a
daily basis (using alcohol or disinfectants, dental practice wipes, such as S
™ wipes).
EPTOL
b) Monitoring
The accessories and cords should be regularly checked for faulty insulation. Replace them if necessary.
It is important to keep the control unit ventilation vents clean in order to avoid abnormal heating.
Check that there is no humidity in the handpiece connector. If there is, dry it out by wiping, then
blowing with the multifunction syringe.
Note:
In the case of a fault, you are advised to contact the supplier of your device. Do not use unauthorized
repairers, who might make your device dangerous for you and your patients.
IX - MAINTENANCE

9. 1 FUSE REPLACEMENT

The device is protected by two fuses located in the mains receptacle (Fig. 2-1).
The replacement procedure is as follows:
- Switch the device off (position O).
- Unplug the power cord from the electric supply.
- Unplug the power cord from the power receptacle (Fig. 2-1).
- Insert the tip of a flat screwdriver into the notch above the fuse drawer to release it (Fig. 2-4).
- Remove the old fuses.
- Replace with new fuses of the same type and rating.
- Push the fuse drawer back into its housing until it clicks to indicate the correct position.
- Plug the power cord into the receptacle (Fig. 2-1).
- Plug the power cord into the electric supply.
Note:
The device also has an internal fuse (ref. FU3 on the printed circuit) which is not accessible to the user.
Contact the SATELEC after-sales service (refer to chapter 2 - Repairs).
All the information necessary for the repair of faulty components is made available by SATELEC to the
technical personnel of the SATELEC-approved dealer network who carry out the work.

9. 2 PUMP REPLACEMENT

Important:
- It is essential to replace a worn pump with a pump supplied by SATELEC.
- In case of difficulty, you can contact the SATELEC after-sales service (see Chapter 2 - Repairs).
The device pump is a "wearing" mechanical part.
To maintain your device in perfect working order, SATELEC advises you to change the pump every year.