VI - USING THE DEVICE
Important:
- Do not disconnect the handpiece from the cord while the machine is plugged in and the footswitch
is pressed.
- For your safety and that of your patient, the device must not be used with accessories other than
those supplied by SATELEC.
- For correct tip vibration the tip must be correctly screwed in, and not forced past its stop point.
To avoid tip self-blockage, it must always be removed after use.
- If it is screwed too far, the tip or handpiece can fracture.
- The irrigation tank cover must always be in place while the device is in use.
- Check the integrity of the device and its accessories before and after each use in order to detect
any problem.
- If necessary, replace all faulty parts.
- Always remove the irrigation tank from the device before filling it.
The device is started as follows:
- Make sure that the device is connected correctly and properly insulated.
- Check that there is no humidity in the handpiece connections or the handpiece cord connector.
Otherwise, dry it out by wiping and blowing with a multifunction syringe.
- Plug the handpiece into the handpiece cord connector.
- Remove the tank from the device.
- Remove the irrigation tank cover.
- Fill the tank with irrigation solution without exceeding the level indicated by the border (Fig. 4-1).
- Replace the cover on the irrigation tank.
- Install the filled tank on its support.
- Set the device to minimum power (1) and maximum flow.
- Start up the irrigation current by pressing the footswitch.
- When the irrigation solution reaches the end of the handpiece, release the footswitch.
- Screw in the desired tip.
- Tighten the tip with moderate force (using the wrench) to ensure optimal ultrasonic operation.
- Set the device to the required power (according to color and number) using the control dial.
- If necessary, refer to the TipBook for ideal SATELEC tip power settings.
- Adjust the irrigation solution flow using the flow adjustment knob.
- Press the footswitch to use the device.
VII - -SHUTTING DOWN THE DEVICE
At the end of the working day, you must:
- Empty the irrigation solution out of the irrigation circuit and the irrigation solution tanks, by pressing
the irrigation flow setting button (Fig. 2, item 3).
- Rinse the circuit and irrigation tank (see chapter IV, 4.4) with distilled or demineralized water.
- Disconnect the handpiece.
- Switch the device off (O).