Problem
Unit does not start on battery
power, although the battery
indicates a charge.
Delay in recharging battery.
Unit alarms, does not start in
car while connected to a
properly functioning DC
socket.
Alarm condition,
Audible alarm sounds
and the (Red) alarm light
illuminates.
Solid (Amber) light without
audible alarm.
All other problems.
®
AirSep
Corporation
MN172-1 rev. - 03/11
Probable Cause
Battery may be too hot
or cold if left outdoors
such as in a car.
Battery exceeds
charging temperature.
Focus power supply
needs resetting.
A general malfunction
has occurred.
Focus need inspection
and / or service.
Solution
Allow the battery to reach
normal operating temperature,
which may take several
minutes if exposed to extremes
of temperature. Temporarily
connect your AC or DC power
supply to the unit's power inlet
and power source, as
necessary.
Unit may be operated;
however, charging may not
resume until battery
temperature is
reduced.
Turn off unit. Disconnect the
DC power supply from the car
socket, restart the car, and
then reconnect the DC power
supply into the car DC socket
to reset the DC power supply.
Turn off unit. Change to
another source of oxygen as
available, and contact your
Equipment Provider.
Contact your Equipment
Provider. It is time for your
Focus unit to be inspected and
/ or serviced by your
Equipment Provider. After any
necessary service and when
the performance is verified by
the Equipment Provider, the
alarm service/indicator light will
be reset.
Turn off unit. Change to
another source of oxygen as
available, and contact your
Equipment Provider.