Charging the Optional AirBelt
To charge the AirBelt battery for extended use:
1)
Release safety cap from end of the AirBelt cord.
2)
Connect the AC/DC power supply (included with AirBelt Accessory kit)
to the end of AirBelt's power cord, as shown in Figure 17.
3)
Connect the AirBelt power supply to an AC electrical socket to recharge.
Figure 17: AirBelt Battery Charging set-up
Do not attempt to charge the optional AirBelt with the Focus
power supply or the AirBelt may be damaged. Use only the
AirBelt power supply provided to charge AirBelt.
The Focus AirBelt will completely recharge from its fully depleted state in
approximately 3½ hours.
While charging a fully discharged battery, the LED will continue to blink until
25% capacity is reached. The LED will then turn solid.
®
AirSep
Corporation
MN172-1 rev. - 03/11
Replace safety cap on AirBelt cord when not in use.