Annex
Guidance and manufacturer´s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and
should only be used in such environments:
IEC 60601
Immunity test
– test
level
Conducted
3 Vrms
RF IEC
150 kHz to
61000-4-6
80 MHz
Radiated
3 V/m
RF IEC
80 MHz to
61000-4-3
2.5 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and refl ection from structures,
objects and people.
Guidance and manufacturer´s declaration – electromagnetic immunity
a)
Field strengths from fi xed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fi xed RF transmitters, an electromagnetic site
survey should be considered. If the measured fi eld strength in the location in which
the device is used exceeds the applicable RF compliance level above, the device
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the device.
b)
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than
3 V/m.
Geratherm Medical AG
Fahrenheitstrasse 1
98716 Geschwenda
Germany
www.geratherm.com
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
3 Vrms
d = 1.2 √P
d = 1.2 √P
3 V/m
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fi xed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the
vicinity ofequipment marked
with the following symbol: