Symptom
No characters on the display.
The display gives erroneous
characters.
The display shows "FIR".
The display shows "
".
The display does not switch
from "
"and "WbC" to three
flashing dashes and "
"
(ready for measuring).
The cuvette holder is not
moving into the correct
position.
Measurement on patient
samples are higher or lower
than antipicipated.
Explanation
1. Power is not received.
2. If on battery power, the batteries need to be
replaced.
3. The display is out of order.
1. The display is out of order.
2. The microprocessor is out of order.
General analyzer software error.
1. The batteries need to be replaced.
2. If on AC power, the power adapter or the
circuit board is out of order.
The cuvette holder sensor is out of order.
The magnet in the cuvette holder is missing.
1. Improper sampling technique.
2. The microcuvettes are beyond their expiration
date, damaged or have been improperly stored.
Action
1a. Check that the power adapter is properly connected to the
analyzer and the AC power supply, as described in the Start Up
section.
1b. Check that the cable is not damaged.
2. Replace the batteries, six type AA, as described in the Start
Up section.
3. The analyzer needs service. Contact your local distributor.
1, 2. The analyzer needs service. Contact your local distributor.
a. Remove and replace all cables and/or batteries, and restart,
as described in the Start Up section.
b. The analyzer needs service. Contact your local distributor.
1. Turn off the analyzer and replace the batteries, six type AA,
as described in the Start Up section.
2a. Check that the correct power adapter is used and properly
connected and working, as described in the Start Up section.
2b. The analyzer needs service. Contact your local distributor.
The analyzer needs service.
Contact your local distributor.
The analyzer needs service.
Contact your local distributor.
1. Take a new microcuvette and repeat the measurement, as
described in the Measuring section.
2. Check the expiration date and the storage conditions of the
microcuvettes.
30